Monitor lead button wire

Model: Trangjan-002
Cable specification
①φ3.5mm electrode buckle
②85A plug material, TPE gray white, 85A plug material, TPE red black, brown green PE 7144
③φ2.0 30/0.08TC pure copper CABLE
④ Lable: Can paste the description information
⑤ ROHS RACEH California 65 environmental protection standard

Category:
Remarks

1. P1 end peeled 15±1mm, P2 peeled 3mm. P3 peeled 3mm
2. P1 P2 P3 all need to be dipped in tin before soldering
3. Use heat shrink tube to protect Soldered DC plug of the grounding at P1 end. P2 and P3 end are directly welded to the related connector.
4. After welding, test T1. Please refer to the previous description for test parameters.
5. Injection molding P1 P2 P3 Test T2 Visual inspection, packaging

Requirements

1. Button SR lifting weight (3kg): ≥ 1 minute.
2. Wire and button holding force: ≥5.0kgf.

Testing requirements

1.a.100% conduction, short circuit, open circuit, dislocation, dielectric withstand voltage test;
b. Insulation resistance: ≥20MΩ, DC 300V, 0.01S;
c. On-resistance: ≤2Ω;
d. Withstand voltage: AC 350V I: 1mA T: 0.1S;
2. Instantaneous short circuit test time ≥ 3 seconds;

Packing

1. Bundle the single cable by PVC mini tie
2. The product is packaged in a PE self-sealing bag with a vent hole, the bag size is 200*300mm, and the thickness is 0.3mm
3. waterproof and dustproof function and packed the products into a large waterproof PE bag of L800MM*W700MM (open end) *T0.04MM transparent color
4. Using medium outer carton K=K
Carton Size: 450MM*320MM*200MM